Title左心室駆出率測定の簡易化, 自動化のための一考案
Subtitle原著
Authors稲垣末次*, 足立晴彦*, 杉原洋樹*, 勝目紘*, 伊地知浜夫*, 岡本邦雄**, 細羽実***
Authors(kana)
Organization*京都府立医科大学第二内科, **RI室, ***島津製作所
Journal核医学
Volume21
Number12
Page1613-1620
Year/Month1984/12
Article原著
Publisher日本核医学会
Abstract「要旨」平衡時心電図同期心プール法による左心室駆出率 (LVEF) 算出において, 関心領域 (ROI) を必要としない新しいバックグランド (BKG) 処理方法を考案し (Fixed BKG ratio法), 心疾患患者および正常者計30名を対象に, 観血的左心室造影法 (LVG) で求めたEFを基準として, 精度, 再現性について, BKG-ROIを用いる従来の方法と比較検討した. BKGを左心室拡張終期カウント数の64%とした時, LVGと最も良く相関 (r = 0.936) するとともに近似値をとり (LVG-EF 61.2±21.4%, Fixed BKG ratio EF 61.1±20.1% (Mean±SD)), 他法よりEF過小評価傾向が是正され, 再現性も良好であった. Fixed BKG ratio法は難解とされるBKG処理の一方法として有用であり, EF算出処理が簡略化されるため, 自動処理への応用やROI設定の困難なシングルプローブシステムへの応用が期待できる.
Practice臨床医学:一般
KeywordsECG gated blood pool scintigraphy, Left ventricular ejection fraction, Background, Radionuclide
English
TitleA New Approach for Simplified and Automated Measurement of Left Ventricular Ejection Fraction by ECG Gated Blood Pool Scintigraphy
SubtitleOriginal Articles
AuthorsSuetsugu INAGAKI*, Haruhiko ADACHI*, Hiroki SUGIHARA*, Hiroshi KATSUME*, Hamao IJICHI*, Kunio OKAMOTO*, Minoru HOSOBA**
Authors(kana)
Organization*Second Department of Internal Medicine, Kyoto Prefectural University of Medicine, **Medical Systems Division, Shimadzu Corporation
JournalThe Japanese Journal of nuclear medicine
Volume21
Number12
Page1613-1620
Year/Month1984/12
ArticleOriginal article
PublisherTHE JAPANESE SOCIETY OF NUCLEAR MEDICINE
Abstract[Summary] Background (BKG) correction is important but debatable in the measurement of Left ventricular ejection fraction (LVEF) with ECG gated blood pool scintigraphy. We devised a new simplified BKG processing (fixed BKG method) without BKG region-of-interest (ROI) assignment, and the accuracy and reproducibility were assessed in 25 patients with various heart diseases and 5 normal subjects by comparison with LVEF obtained by contrast levolgraphy (LVG-EF). Four additional protocols for LVEF measurement with BKG-ROI assignment were also assessed for reference. LVEF calculated using the fixed BKG ratio of 0.64 (BKG count rates were 64%) of end-diastolic count rates of LV) with "Fixed" LV-ROI was best correlated with LVG-EF (r = 0.936, P<0.001) and most approximated (Fixed BKG ratio method EF: 61.1 +- 20.1, LVG-EF: 61.2+-20.4% (mean+-SD)) among other protocols. The wide availability of the fixed value of 0.64 was tested in various diseases, body size and end-diastolic volume by LVG, and the results were to be little influenced by them. Furthermore, fixed BKG method produced lower inter-and intra- observer variability than other protocols requiring BKG-ROI assignment, probably due to its simplified processing. In conclusion, fixed BKG ratio method simplifies the measurement of LVEF, and is feasible for automated processing and single probe system.
PracticeClinical medicine
KeywordsECG gated blood pool scintigraphy, Left ventricular ejection fraction, Background, Radionuclide

【全文PDF】